Site Search:  

Adjustable Coupler for MCT Detector in LN Cooled Dewar

Refocusing coupler for Dewar MCT-detector

Features
  • Designated for detachable IR-fiber optical cables with SMA-termination
  • Adjustable in 3d dimensions - X-Y for lens & Z - for connector position
  • Can be fixed at side window Dewar for QCL / TDL lasers & MCT/ InAs detectors

Standard or customized coupler construction can be supplied for customer laser or detector installed inside LN-cooled Dewar with side window. CIR- or PIR-fiber optical cable with SMA-termination can be supplied according customer specification for his specific applications. 2 types of MCT-detectors are also available for a set with IR-cable and optical fiber coupler - either with LN-cooling in Kedal Dewar for 4-14µm or with triple Peltier cooling - for 1-6µm spectral range.

 

Coupler Construction for LN-cooled detectors or Tunable Diode Lasers

Specification:

Coupler: Ø55 mm, length = 40 mm
Angular translation in the range ±2.50° in both orthogonal axes
Z-translation stage in the range ± 5 mm
Lens: Material - Ge or ZnSe, Ø15 mm, f =10 mm,
AR-coating for 10.6 µm (other wavelength - on request)

 

Coupler Construction for TE-cooled detectors or Quantum Cascade Lasers