Fiber Optic ATR-Probes for Harsh Environment

  • High throughput in selected parts of Near & Mid InfraRed-spectra
  • ATR-tip shaped for immersion in liquid flow without dead zone
  • Robust for industrial applications in harsh environment up to 250°C
  • Resistant to high pressure
  • Compatible with all spectrometers and automated process-interfaces

FlexiSpec® probes from art photonics include new type of Attenuated Total Reflection (ATR-) probes designed for harsh application conditions.
 
HT-ATR immersion IR-fiber probes can be used for process-spectroscopy in Near & Mid IR range to monitor reactions in-line in a broad temperature range from -100° to +250°C. They can resist to high pressures up to 200Bar and used with FTIR or any other IR-spectrometers and spectral sensors in automated process control with process-interfaces.
 
Applications:
  • Remote Reaction Monitoring in-line in temperature range -100/+250°C
  • PAT applications in lab, pilot plant or industry with process-interfaces for automated process control
  • Polymerization Process Control
  • In-situ IR-Spectroscopy for PAT in Chemical, Petrochemical, Atomic, Biopharmaceutical & Food Industry




Specification of IR-Fiber HT-ATR-Probes for High Temperatures FlexiSpec®

Probetype Diamond ATR Silicon ATR Cubic Zirconium ATR
Transmission 5.2-17µm (600-1900cm -1) +
3.2-4.5µm (2300-3100cm-1)
5.2-17µm (600-3100cm -1) +
3.2-4.5µm (2300-3100cm-1)
1.5-6.5µm
(1550-6650cm-1)
Fiber type PIR-900/1000
Silver Halide
PIR-900/1000
Silver Halide
CIR 500/550
Chalcogenide glass (As-S)
Temperature range -100°C / + 250°C -100°C / + 250°C -100°C /+ 200°C
Pressure (max) 200Bar 100Bar 100Bar

Common Parameters of HT-ATR-Probes for High Temperatures FlexiSpec®

Total Length 1.5m (opt.: 1m to 5m)*
Shaft Length 300 mm (opt.: 300-700 mm)*
Shaft Diameter 12mm
Shaft Material Hastelloy C22
Length of Legs 500mm (opt.: 200mm to 500mm)
Protective Tube Material Liquid Tight SS-Conduit, KOPEX-Tube
Minimal Bending Radius 130mm
Input / Output Connectors Long SMA (opt.: any other type)
Cooling Air Flow Parameters Excess pressure 0.5Bar, Flow 2300l/h
Inner temperature control Inside shaft to control ATR tip temperature

Application Notes
Summary Table of ATR Probes
Instruction Manual